12
Ishida Setup & Calibration Manual
2.5
CWS Series
Follow the steps below to set up and calibrate the CWS Series scale.
LB and OZ Calibration Procedure
Step
Action
Display
1
Turn off the scale.
2
Press and hold the TARE and ZERO keys, then press and release the ON/OFF key.
3
Release the TARE and ZERO
after
the display goes blank.
– 1888
4
Press the TARE key while the numbers are scrolling to enter the setup mode. If you do not
press TARE, the CWS proceeds directly to Step 14 for calibration.
1, 2, 3, 4 . . .
5
Enter the scale capacity code using the ZERO key. (See table below).
C1 – N
1
6
Press the TARE key.
C1 – N
1
7
Use the ZERO key to select either: 5 = Auto Off
or
7 = Continuous On.
C2 – N
2
8
Press the TARE key.
C2 – N
2
9
Use the ZERO key to set to F.
NOTE:
Do not enter 6.
C3 – N
3
10
Press the TARE key.
C1 – b
11
Use the ZERO key to set to 4.
C4 – N
4
12
Press the TARE key.
C4 – 2
13
Review the settings. If the settings are correct, press TARE. If not, turn the scale off, then on
again, and re-enter the settings.
2 B N
2
N
1
14
The X’s represent the internal reference counts. Press the ZERO key.
xxxx
15
With no weight on the scale, press ZERO until the display shows 2000 counts.
2000
16
Place the calibration weights on the scale as follows:
5 Lb capacity = 5 Lb weight
10 Lb capacity = 10 Lb weight
19 Lb capacity = 25 Lb weight
2000
17
Press the TARE key.
yyyy
18
The system performs calibration
CAL
19
The span count should read 6000. If not, repeat the setup process from Step 15.
6000
20
Turn the scale over and use a
non-metallic
device to press the memory switch to store the
information (see diagram below).
NOTE:
Using a metallic device can damage the unit. The
display flashes
“P–EP.”
6000
21
Return the scale to an upright position, then turn off the scale.
6000
22
Turn the scale on again and test the calibration by placing a known weight on the scale.
CWS-oz-05
CWS-oz-10
CWS-oz-19
Capacity
Code
Capacity
Code
Capacity
Code
5 Lb x 0.1
oz
0
10 lb x 0.2 oz
1
19 lb x 0.5 oz
2
Table 2-2. Lb/Oz Scale capacity codes
Memory Switch
Содержание BC-3000
Страница 1: ...51860 Ishida Industrial Non System Retail Products Setup Calibration Manual ...
Страница 2: ......
Страница 4: ......
Страница 8: ...4 Ishida Setup Calibration Manual ...