16
Ishida Setup & Calibration Manual
3.1
QB Series
Follow the steps in the table below to set up and calibrate the QB Series scale.
NOTES:
•
Press the
P
key to cancel calibration procedures and return to the weigh display.
•
If
[o–Err]
appears on the display, the weight exceeds the scale capacity.
•
Standard calibration weight is
1/2
capacity. However, when possible, calibrate using a full capacity
weight. See the table below for the unit system capacities.
•
If
[1–Err]
appears on the display, the calibration weight is less than 50% of the standard.
•
If
[2–Err
] appears on the display, an object other than the calibration weight is on the pan, or the display
tolerance (1%) is exceeded.
Step
Action
1
Press and hold the F key until [Func] –> [CAL] appear in sequence, then release the key.
2
Press and hold the T key, then press the F key, then release both keys.
3
[un it] appears. Press the T key to increase the value of the right digit in the following order:
[0] –> [1] –> [2] –> . . . .[9] –> [A] –> [6] –> [d] (repeat from 0).
Select the desired unit system number (see table below).
4
Press the F key to call up [on 0]. The QB automatically performs a zero adjustment. (Do not leave any weight on
the pan).
5
After the QB performs zero adjustment, [on FS] automatically appears. Gently place the designated calibration
weight on the center weigh pan (see table below for designated calibration weights).
6
When calibration is complete, the scale automatically returns to the weighing mode.
Unit No. (mark)
QB-150E
QB-300E
QB-600E QB-1500E QB-3000E QB-6000E QB-12KE
QB-620E QB-6200E
1 (g)
150
300
600
1500
3000
6000
12000
620
6200
2 (kg)
0.15
0.3
0.6
1.5
3
6
12
0.62
6.2
3 (ct)
750
1500
3000
7500
15000
30000
60000
3100
31000
4 (oz)
5.2
10
21
52
100
210
420
21
210
5 (Lb)
0.33
0.66
1.3
3.3
6.6
13
26
1.3
13
6 (ozt)
4.8
9.6
19
48
96
190
380
19
190
7 (dwt)
96
190
380
960
1900
3800
7700
390
3900
8 (gr)
2300
4600
9200
23000
46000
92000
180000
9500
95000
9 (tl)
4
8
16
40
80
160
320
16
160
A (tl)
3.9
7.9
15
39
79
150
310
16
160
b (tl)
4
8
16
40
80
160
320
16
160
c (mom)
40
80
160
400
800
1600
3200
160
1600