14
Ishida Setup & Calibration Manual
2.7
QBX Series
Follow the steps below to set up and calibrate the QBX Series scale.
NOTES:
•
Press any key to cancel calibration procedures and return to the weigh display.
•
If
[o–Err]
appears on the display, the weight exceeds the scale capacity.
•
Standard calibration weight is
1/2
capacity. However, when possible, calibrate using a full capacity
weight. See the table below for the unit system capacities.
•
If
[1–Err]
appears on the display, the calibration weight is less than 50% of the standard.
•
If
[2–Err]
appears on the display, an object other than the calibration weight is on the pan, or the display
tolerance (1%) is exceeded.
Setup
Action
1
Press and hold the MODE key until [Func] –> [CAL] appear in sequence, then release the key.
2
Press and hold the T key, press the MODE key, then release both keys at the same time.
3
[un it] appears. Press the T key to increase the value of the right digit in this order:
[1] –> [2] –>. . . . [5]
Select the desired until system number (see table below).
4
Press the MODE key to call up [on 0]. The QB automatically performs zero adjustment.
(Do not leave any weight on the pan).
5
After the QBX performs zero adjustment, [on FS] is shown. Place the calibration weight on the center of the
weigh pan (see table below for designated calibration weights).
6
When calibration is complete, the scale automatically returns to weighing mode.
Unit No. (mark)
QB–150E
QB–300E
QB–600E QB–1500E QB–3000E QB–6000E QB–12KE
QB–620E QB–6200E
1 (g)
150
300
600
1500
3000
6000
12000
620
6200
2 (kg)
0.15
0.3
0.6
1.5
3
12
12
0.62
6.2
3 (ct)
750
1500
3000
7500
15000
30000
60000
3100
31000
4 (oz)
5.2
10
21
52
100
210
420
21
210
5 (lb)
0.33
0.66
1.3
3.3
6.6
13
26
1.3
13
Содержание BC-3000
Страница 1: ...51860 Ishida Industrial Non System Retail Products Setup Calibration Manual ...
Страница 2: ......
Страница 4: ......
Страница 8: ...4 Ishida Setup Calibration Manual ...