2.4
CWS Series
Follow the steps below to set up and calibrate the CWS Series scale.
KG and GRAM Calibration Procedure:
Step
Action
Display
1
Turn off the scale.
2
Press and hold the TARE and ZERO keys, then press and release the ON/OFF key.
3
Release the TARE and ZERO
after
the display goes blank.
–1888
4
Press the TARE key while the numbers are scrolling to enter the setup mode. If you do not
press the TARE, the CWS proceeds directly to Step 14 for calibration.
1, 2, 3, 4. . .
5
Enter the scale capacity code using the ZERO key. (see table below).
C1 – N
1
6
Press the TARE key.
C1 – N
1
7
Use the ZERO key to select either: 5 = Auto Off
or
7 = Continuous On.
C2 – N
2
8
Press the TARE key.
C2 – N
2
9
Use the ZERO key to set b.
NOTE:
Do not enter 6.
C3 – N
3
10
Press the TARE key.
C3 – b
11
Use the ZERO key to set 2.
C4 – N
4
12
Press the TARE key.
C4 – 2
13
Review the settings. If the settings are correct, press TARE. If not, turn the scale off, then on
again, and re-enter the settings.
2 B N
2
N
1
14
The X’s represent the internal reference counts. Press the ZERO key.
xxxx
15
With no weight on the scale, press ZERO until the display shows 2000 counts.
2000
16
Place the full capacity weights on the scale as determined in Step 5 above.
2000
17
Press the TARE key.
yyyy
18
The system performs calibration.
CAL
19
The span count should read 7000. If not, repeat the setup process from Step 5.
7000
20
Turn the scale over and use a
non-metallic
device to press the memory switch to store the
information (see diagram below).
NOTE:
Using a metallic device can damage the unit. The
display flashes “P-EP”.
7000
21
Return the scale to an upright position, then turn off the scale.
7000
22
Turn the scale on again and test the calibration by placing a known weight on the scale.
CWS–02
CWS–05
CWS–10
CWS–20
Capacity
Code
Capacity
Code
Capacity
Code
Capacity
Code
2000 x 2 g
4
5000 x 5 g
8
10 x 0.01 kg
2
20 x 0.02 kg
6
2 x 0.002 kg
7
5 x 0.005 kg
B
20 x 0.02 lb
6
50 x 0.05 lb
A
5 x 0.005 lb
B
10 x 0.01 lb
2
100 x 0.1 oz
1
200 x 0.2 oz
5
Table 2-1. Scale capacity codes