1.8 Connector assignment
1.8.1 Power connection
The unit is connected to mains net using the power connector on the back panel.
1.8.2 High voltage connection
DANGER!
DANGER!
Before operations at the load or the high voltage output of the power supply are started, the device has to be
switched off, the discharge of residual voltage has to be finished and the high voltage output of the power
supply must be properly grounded. Depending on application residiual voltages can be present for long time
periods. These residiual voltages can lead to severe injuries.
The device has one high voltage output.
The HV cable has to be connected to the load properly and isolated according to proof-voltage.
The shield of the high voltage output is always connected to the housing and can be used as a return conductor.
For the GPS 300W and 800W device classes, the factory installed short-circuit bridge (see Figure 10: Example illustration of the
short-circuit bridge) must be mounted between the connections "0V" and "
".
Figure 10: Example illustration of the short-circuit bridge
Devices of the 350W - COMPACT series have an ground bolt (M4 thread).
Figure 11: ground bolt