WARNING!
WARNING!
Do not operate the unit if you suspect the unit or the connected equipment to be damaged.
WARNING!
WARNING!
The mains connection is made with basic insulation and protective conductor. The device may only be operated
with the protective earth conductor (PE) connected!
The protective conductor connections must be checked for proper function after installation.
WARNING!
WARNING!
The protective conductor connection must be ensured by an appropriate mains cable. Before connecting to the
local power supply, check whether the nominal voltage of the devices corresponds to the mains voltage.
WARNING!
WARNING!
For the device classes GPS, 300W and 800W, the user must ensure that the potential difference between the
return conductor and the protective conductor does not pose a hazard!
For the device classes GPS, 300W and 800W, the interface connection has the same potential as the "0V"
connection.
WARNING!
WARNING!
For the device classes GPS 300W and 800W the factory installed short-circuit bridge must be mounted between
the connections "0V" and "
".
If this short-circuit bridge is not installed, an additional conductor with a cross-section of at least 1.5 mm² must
be used as the return conductor. This is connected to the "0V" connection.
The "0V" terminal can accept a voltage related to the earth potential.
WARNING!
WARNING!
At an ambient temperature greater than 35°C, the temperature of the housing can exceed 45°C!