S p e z i a l e l e k t r o n i k G m b H
iseg
c/o ROTECH
Email: [email protected]
Phone ++ 49 351 / 26 95 - 260
Spezialelektronik
Bautzner Landstr. 45
http://www.iseg-hv.com
Fax
++ 49 351 / 26 95 - 261
GmbH
D - 01454 Rossendorf
Germany
13
Group access : Channel works according to control
(Read-write/Write access),
extended access list
Read-write
Byte
Identifier
DATA_ID
Controller (DLC = 1):
Bit
ID1
ID0
7
6
5
4
3
2
1
0
Are channels working correctly according to
Designation
EXT_
INSTR
DATA
_DIR
control?
Data
1
1
1
1
0
0
1
1
0
0
Description
read
⇓
Response module (DLC = 3)
Byte
Identifier
DATA_ID
DATA_1
DATA_0
Bit
ID1
ID0
7
6
5
4
3
2
1
0
7
...
0
7
...
0
Designation
EXT_
INSTR
DATA
_DIR
N3 N2 N1 N0
LSB
Data
1
0
1
1
0
0
1
1
0
0
x
Description
write
x=1: Channel is working correctly
x=0: Channel is not working correctly
LSB = Channel 1
Group access : Status4
(Read-write/Write access),
extended access list
Read-write
Byte
Identifier
DATA_ID
Controller (DLC = 1):
Bit
ID1
ID0
7
6
5
4
3
2
1
0
Control of sense line (is Vset = Vsense ?)
Designation
EXT_
INSTR
DATA
_DIR
Data
1
1
1
1
0
1
0
0
0
0
Description
read
⇓
Response module (DLC = 3)
Byte
Identifier
DATA_ID
DATA_1
DATA_0
Bit
ID1
ID0
7
6
5
4
3
2
1
0
7
...
0
7
...
0
Designation
EXT_
INSTR
DATA
_DIR
N3 N2 N1 N0
LSB
Data
1
0
1
1
0
1
0
0
0
0
x
Description
write
x=1: Vset
≠
Vsense (Sense error)
x=0: Vset = Vsense (
±
tolerance)
LSB = Channel 1