26
INDUSTRIAL SCIENTIFIC
Revision 5 (p/n: 17120932)
Set Display Type
The Set Display Type screen is used to set the display type for monitoring in
normal instrument operation. Pressing the Enter button on this screen will toggle
between the two display modes – Gas Reading, or Sensor Type Only. On the Gas
Reading screen, the actual current gas concentration will be placed on the primary
display, and the sensor type on the auxiliary display, along with the up and down
arrows to indicate that the display type can be modified. On the Sensor Type Only
screen, the sensor type will move to the primary display, and the auxiliary display
will be blank. Pressing Mode will move to the Set Confidence Indicator screen.
Set Display Type - Gas Reading (Left) and Sensor Type Only (Right)
Actions Available from the Set Display Type Screen
Action Response
Mode button press
Go to Set Confidence Indicator
Enter button press
Toggle display type setting
Mode and Enter buttons
held (3 sec.)
+
Go to Gas Monitoring screen
30 second timeout
Go to Gas Monitoring screen