30
INDUSTRIAL SCIENTIFIC
Revision 5 (p/n: 17120932)
Instrument Test Screen
On this screen, you can initiate an
instrument test by pressing the Enter
button. The instrument test lights all
segments of the display for five
seconds, and runs an indicator test
(similar to the test that is performed at
instrument startup). Pressing the Mode
button jumps back to the first screen in
the Configuration Mode loop.
Instrument Test Screen
Actions Available from the Instrument Test Screen
Action Response
Mode button press
Go to Zero Initiate screen
Enter button press
Go to Test Flash screen (all segments are
turned on and each alarm flashes). This
screen will time out back to the
Instrument Test screen.
Mode and Enter
buttons held
+
Go to Gas Monitoring screen
30 second timeout
Go to Gas Monitoring screen
Bump in Field Option
This screen allows the user to enable the option to
perform a bump test in the field. The default
setting on this option is 0. When this screen is set
to 1, it will allow the “bump” screen to be seen in
the normal operating mode. Setting this screen to 1
will also enable further configuration options for
enabling the bump due alarm, the bump test
interval, and the bump test duration.