Revision 5 (p/n: 17120932)

        INDUSTRIAL SCIENTIFIC                                                   

 

17 

Actions Available from the Zero Initiate Screen 

Action Response 

Mode button press 

Go to Days Since Calibration screen

Enter button press 

Go to Zero In Process 

Mode and Enter buttons 
held (3 sec.) 

+

Go to Gas Monitoring screen 

30 second timeout 

Go to Gas Monitoring screen 

 

Zeroing Failed 

The Zeroing Failed screen is displayed as a 
zero icon and a flashing warning icon.  The 
instrument will remain in periodic alarm (one 
alarm burst every 15 seconds) while in this 
condition, to attempt to conserve battery life. 

From this screen you only have the option to 
repeat the zeroing process by pressing the 
Enter button.

 

 

Zeroing Failed Screen 

 

NOTE:

  If the instrument is connected to the Cal Plus

 calibration station while in 

Zero Fail state, the alarms will NOT activate until the instrument is disconnected 
from the bump/calibration station. 

Actions Available from the Zeroing Failed Screen 

Action Response 

Mode button press 

No response 

Enter button press 

Go to Zeroing In Process 

Mode and Enter buttons 
held (3 sec.) 

+

No response 

 
 

Содержание GasBadge Plus 3.0

Страница 1: ...e Gas Monitor V3 0 Activation Operation Troubleshooting Part Number 17120932 1 Revision 5 Ex Ox Tox Gasdetectie Westerdreef 5V 2152 CS Nieuw Vennep Telefoon 0252 620885 E mail info exoxtox nl Website...

Страница 2: ...n 18 Days Since Calibration 21 Event Log 21 Initiate Print 22 Low Alarm Setpoint 23 High Alarm Setpoint 24 Cal Gas Setpoint 25 Set Display Type 26 Set Confidence Indicator 27 Set Security Code 28 Tota...

Страница 3: ...ended by Directives 92 31 EEC and 92 31 EEC and 93 68 EEC The EC Type Examination Certificate is DEMKO 09 ATEX 0844098 with marking Code Ex ia IMa Ex ia IIC T4 Ga for Equipment Group and Category I M1...

Страница 4: ...1 17124033 Cal Cup 1 17120643 Replacement Sensor Water Barrier Filter 1 17121385 Quick Start Guide After unpacking if any listed item is missing contact either your local distributor of Industrial Sci...

Страница 5: ...down GasBadge Pro Only GasBadge Pro Only GasBadge Pro Only Alarm Auxiliary Display Confidence and Warning Icons Peak Value Zeroing Battery Status Apply Gas Calibration Security Main Display Display T...

Страница 6: ...he instrument will display the software version IMPORTANT Once the GasBadge Plus is activated it cannot be turned off NOTE Any button press activates the backlight for a period of 3 seconds Countdown...

Страница 7: ...erating Quick Start Guides Basic Operating Flowchart 1 Minute Months Days Hours Gas Reading Screen Battery Life Remaining Screen Peak Reading Screen Bump Test Screen if enabled Hold 3 seconds to enter...

Страница 8: ...bration Print Event Log Low Alarm Setpoint Adjust High Alarm Setpoint Adjust Calibration Gas Value Adjust Display Setting Adjust Confidence Beep Enable Security Code Setting Total Instrument Alarm Tim...

Страница 9: ...ition back to the Gas Monitoring screen The alarm screen is indicated on the display by the Alarm Indicator and either the Up Indicator or Down Indicator which designate a low or high alarm High Alarm...

Страница 10: ...en as low medium or full The Month Icon will be active to indicate that this is the number of months The instrument will continue to indicate the sensor type on the auxiliary display and the Checkmark...

Страница 11: ...screen When the instrument reaches the end of its useful life the word FAIL will appear on the instrument s display and both the audible and visual alarms will be activated the unit should be removed...

Страница 12: ...the Enter button on the Peak Gas Reading screen will clear the peak value On O2 instruments the depletion peak will be cleared to 20 9 NOTE The GasBadge Plus retains the maximum gas reading for a tox...

Страница 13: ...the Bump Test Screen Action Response Enter button press Initiate bump test Mode button press Go to Peak Reading screen 30 second timeout Go to Gas Monitoring screen If the sensor responds to 50 of th...

Страница 14: ...mp test using a known concentration of calibration gas should be performed periodically based on instrument use exposure to gas and environmental conditions The frequency is best determined by company...

Страница 15: ...ment will return to the Gas Monitoring screen if no buttons are pressed for 30 seconds Security Code The Security Code screen is the first screen in the configuration mode The default security code is...

Страница 16: ...PORTANT Before zeroing ensure the instrument is located in a fresh air environment If background gas is present the use of a zero air cylinder is recommended The Zero screen allows you to begin the ze...

Страница 17: ...will remain in periodic alarm one alarm burst every 15 seconds while in this condition to attempt to conserve battery life From this screen you only have the option to repeat the zeroing process by pr...

Страница 18: ...Enter button press Go to Calibration screen Mode and Enter buttons held 3 sec Go to Gas Monitoring screen 5 second timeout Go to Calibration screen Calibration Calibrating with Ambient Air Oxygen Sens...

Страница 19: ...with the supplied tubing On the Calibration screen the Cal Gas bottle will flash notifying you to apply the gas concentration shown and press Enter to begin calibration Calibrate using a flow rate of...

Страница 20: ...ays Since Last Cal screen Enter button press No response No response Mode and Enter buttons held 3 sec Go to Gas Monitoring screen Go to Gas Monitoring screen Actions Available from the Cal Failed Scr...

Страница 21: ...sec Go to Gas Monitoring screen 30 second timeout Go to Gas Monitoring screen Event Log The GasBadge Plus stores alarm events in non volatile memory The last 15 gas alarm events are stored with conti...

Страница 22: ...ed from the instrument the Print icon will blink Initiate Print Screen Actions Available from the Initiate Print Screen Action Response Mode button press Go to Low Alarm Setpoint screen Enter button p...

Страница 23: ...y pressing the Enter button Simple momentary presses will increment the value by 1 unit Holding the Enter button for an extended time will increment the value much faster Low Alarm Setpoint Screen Upo...

Страница 24: ...lding the Enter button for an extended time increments the value much faster High Alarm Setpoint Screen Upon reaching the maximum value which is sensor dependent the display rolls over to the minimum...

Страница 25: ...are ready to change The threshold is incremented by pressing the Enter button Simple momentary presses will increment the value by 1 unit Cal Gas Setpoint Holding the Enter button for an extended tim...

Страница 26: ...uxiliary display along with the up and down arrows to indicate that the display type can be modified On the Sensor Type Only screen the sensor type will move to the primary display and the auxiliary d...

Страница 27: ...ge The Enter button will toggle the rightmost character between 0 and 1 The Mode button will cause a transition to the Set Security Code screen Set Confidence Indication Actions Available from the Set...

Страница 28: ...e Total Alarm Time screen Set Security Code Screen Actions Available from the Set Security Code Screen Action Response Mode button press Accept current code then go to the Total Alarm Time screen Ente...

Страница 29: ...Version On this screen the software version and revision are displayed The auxiliary display will indicate the software build number as an integer from 0 to 99 preceded by the letter b Pressing Mode t...

Страница 30: ...utton press Go to Zero Initiate screen Enter button press Go to Test Flash screen all segments are turned on and each alarm flashes This screen will time out back to the Instrument Test screen Mode an...

Страница 31: ...arm Enable If the bump test option has been enabled this screen allows the user to enable an alarm that indicates when the instrument is overdue for bump testing If the bump overdue alarm is enabled t...

Страница 32: ...en TCM Action Response Mode button press Go to Bump Test Timeout screen Enter button press Increase value from 0 5 to 7 0 in 0 5 increments Default is 1 0 Bump Test interval is one day 30 second timeo...

Страница 33: ...fter the USB drivers are installed the Datalink user interface software can be installed Once installed and running the Datalink software allows users to communicate to either the Datalink Accessory o...

Страница 34: ...Station provides the ultimate flexibility for managing gas monitors wherever they are used Ethernet connectivity enables the linking of up to 700 stand alone Instrument Docking Stations IDSs from remo...

Страница 35: ...the unit should be removed from service immediately The unit will remain in this alarm state for 10 minutes during this time it will no longer detect gas After 10 minutes in this alarm state the word...

Страница 36: ...xpected Instrument Error The Unexpected Instrument Error screen indicates that an unexpected instrument error has occurred and the instrument will reset If the error persists the instrument must be re...

Страница 37: ...est may indicate that the membrane is blocked An obstructed filter membrane must be replaced prior to further use of the gas monitor To replace the membrane Remove the obstructed membrane from the ins...

Страница 38: ...106419 2 unit nylon carrying case 17124033 Calibration cup Cal Cup 18106260 GasBadge Datalink 18106344 0X2 Cal Plus Calibration Station 18106344 1X2 Cal Plus Calibration Station with on board dot matr...

Страница 39: ...graphical icons for easy use Segmented display for direct gas readings Backlight for low light conditions User selectable option for direct gas reading mode or display just gas type Peak reading indic...

Страница 40: ...0 4 to 122 0 99 SO2 0 150 ppm 0 1 ppm 5 0 15 40 to 50 40 to 122 15 90 1 For any given sensor reading the accuracy shall be taken to be the greater value of the value specified below or 1 count 2 Withi...

Страница 41: ...or in tort including strict liability in tort It shall be an express condition to Industrial Scientific s warranty that all products be carefully inspected for damage by Buyer upon receipt be properl...

Страница 42: ...42 INDUSTRIAL SCIENTIFIC Revision 5 p n 17120932 Ex Ox Tox Gasdetectie Westerdreef 5V 2152 CS Nieuw Vennep Telefoon 0252 620885 E mail info exoxtox nl Website www exoxtox nl...

Страница 43: ...Revision 5 p n 17120932 INDUSTRIAL SCIENTIFIC 43...

Отзывы: