Revision 5 (p/n: 17120932)
INDUSTRIAL SCIENTIFIC
17
Actions Available from the Zero Initiate Screen
Action Response
Mode button press
Go to Days Since Calibration screen
Enter button press
Go to Zero In Process
Mode and Enter buttons
held (3 sec.)
+
Go to Gas Monitoring screen
30 second timeout
Go to Gas Monitoring screen
Zeroing Failed
The Zeroing Failed screen is displayed as a
zero icon and a flashing warning icon. The
instrument will remain in periodic alarm (one
alarm burst every 15 seconds) while in this
condition, to attempt to conserve battery life.
From this screen you only have the option to
repeat the zeroing process by pressing the
Enter button.
Zeroing Failed Screen
NOTE:
If the instrument is connected to the Cal Plus
calibration station while in
Zero Fail state, the alarms will NOT activate until the instrument is disconnected
from the bump/calibration station.
Actions Available from the Zeroing Failed Screen
Action Response
Mode button press
No response
Enter button press
Go to Zeroing In Process
Mode and Enter buttons
held (3 sec.)
+
No response