Zero Passed
The Zero Passed screen shows that the
zero process completed successfully.
After 5 seconds, the Calibration screen
is displayed.
Zero Passed Screen
Actions Available from the Zero Passed Screen
Action Response
Mode button press
Go to Zero Initiate screen
Enter button press
Go to Calibration screen
Mode and Enter buttons
held (3 sec.)
+
Go to Gas Monitoring screen
5 second timeout
Go to Calibration screen
Calibration
Calibrating with Ambient Air
(Oxygen Sensors Only)
NOTE:
Clean ambient air may be used
for oxygen calibration. If a clean air
environment is in question or if a zero air
cylinder calibration is preferred, please
skip to the Calibrating with Gas
Cylinders section.
On the Calibration screen, the Cal Gas
bottle will flash, notifying you to press
Calibration Cup