PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems
Page 79 of 141
88292-EN, Rev D
Pattern Titration
A
<Titrate>— press to switch to Titrate Mode to perform test lesions in a linear pattern,
adjustable from 1 to 4 spots; press again to revert to prior laser mode
B
Power Display – Maximum and Minimum are displayed when 2 or more
applications are selected.
C
<Spacing>—1.0 to 2.0 diameters
D
Titration Pattern—press to select number of titration applications
<Titrate> provides the option to deliver 1 to 4 laser applications in a linear pattern to facilitate
selection of desired laser dosimetry. Power within the titration pattern decreases from left to
right, starting with the full power setting and decreasing by one power increment for each spot,
moving to the right across the pattern. The titration increment will vary depending on the full
power setting and corresponds to the incremental step change in power shown by adjusting
the power down arrow, e.g. a 4-spot titration pattern starting at 175 mW will deliver
175mW
à
150mW
à
140mW
à
130mW.
WARNING
Please confirm that Endpoint MGMT is automatically disabled and grayed out during titration
mode.
Performing titration when EndPoint Management is turned on may result in overpower
exposure.