Calibration Procedure
Regulatory agencies require that manufacturers of US FDA CDRH Class II and IV and
European IEC 60825 Class 2 and 4 medical lasers supply customers with power calibration
instructions.
Calibration must be performed by an engineer or technician qualified to work on energized
electronic laser equipment.
Disclaimer Warning
Calibration of the system is a service procedure to be performed only by certified personnel.
Adjustment by anyone other than certified personnel voids any existing manufacturer’s
warranty on the instrument and can result in serious personal injury.
Calibration Instructions
There is no need to open the system for calibration. In fact calibration should be done
with the covers on.
Tools Required:
•
USB keyboard and mouse
•
NIST-traceable Gentec calibrated optical power meter
•
Power meter to chinrest adapter
1. De-energize the system.
2. Connect the keyboard and mouse to the USB port.
3. Attach the power meter to chinrest adapter.
4. Install the optical power meter.
5. Re-energize and start up the system.
6. Navigate to the Service/Calibration software and enter Service Mode.
7. Align the optical power meter using the aiming beam.
8. Calibrate the system as per DC-03366 service manual.
9. Shut down and de-energize the system, as indicated in the <System
Start-up and Shut-down> section in this manual.
10. Remove the keyboard and mouse.
11. Re-energize and start up the system.
12. Verify the calibration using the optical power meter.
If successful, shut down the system and remove the power meter and chinrest adapter. If not
successful, repeat the calibration procedure.