PASCAL
®
Synthesis Ophthalmic Scanning Laser Systems
Page 117 of 141
88292-EN, Rev D
Code
Message
Critical
Clearable
Warning
Description
Action
61
Red 60µm Treatment
Local Light Under
(Low)
X
Laser output is 20% under the
expected value
User to select End
Treatment and Return to
the Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats- End
Treatment and call Service
62
Red 60µm Treatment
Local Light Over
(Low)
X
Laser output is 20% over the
expected value
User to select End
Treatment and Return to
the Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats- End
Treatment and call Service
63
Red 60µm Treatment
Local Light Over
(Medium)
X
Laser output is 50% over the
expected value
User to select End
Treatment and Return to
the Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats- End
Treatment and call Service
64
Red 60µm Treatment
Local Light Over
(High)
X
Laser output is 100% over the
expected value
User to select End
Treatment and Return to
the Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats- End
Treatment and call Service
65
Red 200µm
Treatment Local
Light Under (Low)
X
Laser output is 20% under the
expected value
User to select End
Treatment and Return to
the Home Screen.
Re-access treatment
screen and reconfigure
treatment parameters.
If error repeats- End
Treatment and call Service
66
Red 200µm
Treatment Local
Light Over (Low)
X
Laser output is 20% over the
expected value
User to select End
Treatment and Return to
the Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats- End
Treatment and call Service
67
Red 200µm
Treatment Local
Light Over (Medium)
X
Laser output is 50% over the
expected value
User to select End
Treatment and Return to
the Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats- End
Treatment and call Service
68
Red 200µm
Treatment Local
Light Over (High)
X
Laser output is 100% over the
expected value
User to select End
Treatment and Return to
the Home Screen.