PASCAL
®
Synthesis TwinStar Ophthalmic Scanning Laser Systems
Page 112 of 142
88293-EN, Rev C
Error Messages
Error States
If error occurs regularly, contact service.
Critical Error
implies a safety issue that requires the system to immediately go to a safe
state. System restart may be required.
Clearable Error
indicates a problem with the system that requires that the system be put
into a safe state until the error is acknowledged by the user.
Warning
indicates a problem with the system that does not represent a serious safety
issue and does not require any system functions to be interrupted, but does require that
the user be notified.
Code
Message
Critical
Clearable
Warning
Description
Action
01
Kill Line
X
The kill lines were asserted in the
hardware
Reboot.
Call service if error
persists.
02
Emergency Stop
X
The emergency stop switch was
pressed.
Release Emergency Stop.
03
Footswitch Connect
X
Footswitch has malfunctioned or
been disconnected
Connect Footswitch
04
Power Rails
X
An error was detected with the
power supplies.
Reboot.
Call service if error persists.
05
Watchdog (Galvo)
X
The watchdog fired; this implies
the back-end is taking too
long in the ISR, or is in an infinite
loop.
Reboot.
Call service if error persists.
06
Watchdog (Main)
X
The watchdog fired; this implies
the back-end is taking too
long in the ISR, or is in an infinite
loop.
Reboot.
Call service if error persists.
07
Interlock
X
The interlock switch
was activated.
Connect Remote Interlock
11
Aiming Current Over
(High)
X
Aiming beam output above
threshold.
Call service if error persists.
15
OPSL Treatment
Current Over
(Medium)
X
OPSL laser draws electric current
above threshold.
Call service if error persists.
16
OPSL Treatment
Current Over (High)
X
OPSL laser draws electric current
above threshold.
Call service if error persists.
17
Unexpected OPSL
Treatment Current
X
Electric current was drawn
by OPSL Laser module
when it was not expected
Verify device connections.