PASCAL
®
Synthesis TwinStar Ophthalmic Scanning Laser Systems
Page 19 of 142
88293-EN, Rev C
An eye safety filter is provided with the Slit Lamp and is required for safe use. Laser safety
eyewear is not required by the physician who views the procedure through the slit lamp
eyepieces. All other personnel within the NHZ must wear laser safety eyewear with the
recommended optical density.
Along with providing the appropriate safety eyewear, the following steps should be taken
to secure the controlled area:
1. Treatment should be conducted in a dedicated, enclosed room.
2. A warning sign should be placed on the outside of the treatment room door when the
laser is in use. The sign is intended to alert personnel before they enter the controlled
area.
3. The treatment room door should be kept closed during treatment.
Electrical Hazards
WARNING: DANGEROUS VOLTAGE
To avoid the risk of electric shock, this equipment must only be connected to a supply
mains with protective earth. Hospital Grade Cord grounding reliability can only be
achieved when equipment is connected to an equivalent receptacle marked “Hospital
Grade” or “Hospital only”.
To avoid the risk of electric shock, do not touch any external connector and the patient
simultaneously.
Do not use power cables other than the power cable provided with the system. Do not use
extension cables with the system.
Disconnect the laser system from the electrical outlet when inspecting the fuses.
Never open the laser console protective covers. Opening the covers will expose you to
high voltage components, the laser resonator and possible laser radiation. Only certified
personnel shall work inside the console.
The area around the laser and footswitch shall be kept dry. Do not operate the laser if any
of the cords are faulty or frayed. The laser should undergo routine inspection and
maintenance per the Iridex Corporation manufacturer's recommendation and institutional
standards.
Non Homogeneous Pattern Delivery
WARNING
The optical system in this unit has been designed to provide uniform laser energy
deposition over the entire area of a focused spot. If proper focus of the laser spot on its
intended target is not achieved, or if damage, contamination, or deterioration of the optical
components has occurred, it is also possible to degrade this uniformity. Observation of
Содержание PASCAL Synthesis TwinStar