Pattern Titration
A
<Titrate>—press to switch to Titrate Mode to perform test lesions in a linear pattern,
adjustable from 1 to 4 spots; press again to revert to prior laser mode
B
Power Display – Maximum and Minimum are displayed when 2 more applications are
selected.
C
<Spacing>—1.0 to 2.0 diameters
D
Titration Pattern - press to select number of titration applications
<Titrate> provides the option to deliver 1 to 4 laser applications in a linear pattern to
facilitate selection of desired laser dosimetry. Power within the titration pattern decreases
from left to right, starting with the full power setting and decreasing by one power
increment for each spot, moving to the right across the pattern. The titration increment will
vary depending on the full power setting and corresponds to the incremental step change
in power shown by adjusting the power down arrow, e.g. a 4-spot titration pattern starting
at 175 mW will deliver 175mW
à
150mW
à
140mW
à
130mW.
WARNING
Please confirm that Endpoint MGMT is automatically disabled and grayed out during
titration mode .
Performing titration when EndPoint Management is turned on may result in overpower
exposure.
Refer to Warning Section for Pattern Titration Warning Information.