PASCAL
®
Synthesis TwinStar Ophthalmic Scanning Laser Systems
Page 116 of 142
88293-EN, Rev C
Code
Message
Critical
Clearable
Warning
Description
Action
52
Ambient Temperature
Over Maximum
X
Ambient Temperature during
treatment exceeds maximum
allowable value
Allow system to cool.
Reconfigure treatment
parameters if necessary
56
Aiming Local Light
Over (High)
X
Aiming laser output is 100% over
the expected value
User to select End
Treatment and Return
to the Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service
57
OPSL Treatment
Local Light Under
(Low)
X
Laser output is 20% under the
expected value
User to select End
Treatment and Return
to the Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service
58
OPSL Treatment
Local Light Over (Low)
X
Laser output is 20% over the
expected value
User to select End
Treatment and Return
to the Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service
59
OPSL Treatment
Local Light Over
(Medium)
X
Laser output is 50% over the
expected value
User to select End
Treatment and Return to the
Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service
60
OPSL Treatment
Local Light Over
(High)
X
Laser output is 100% over the
expected value
User to select End
Treatment and
Return to the Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service
61
Red 60µm Treatment
Local Light Under
(Low)
X
Laser output is 20% under the
expected value
User to select End
Treatment and Return to the
Home Screen.
Re-access treatment screen
and reconfigure treatment
parameters.
If error repeats-End
Treatment and call Service