Primary pattern
Secondary pattern
Array
General Use
•
retina tears, retina detachments
Pattern
Spot Diameter (μm)
Spacing
Array
50 (2x2 or smaller),
0.00Ø to 3.00Ø
100
200
400
0.00Ø to 1.50Ø
The Array (20 ms default) pattern is selectable in a variety of shapes and sizes up to a
maximum of 9 spots, including square arrays, rectangle arrays, vertical and horizontal
lines of up to 3 spots, and a single spot. Spot diameter and spacing settings are also
adjustable.
To select the Array pattern shape and size, drag your finger horizontally, vertically, or
diagonally across the pattern, or press one of the quick-select buttons at the bottom of the
secondary pattern screen. To select the pattern orientation, press <Rotate> button at the
bottom of the screen.