Invacare SB 400 Скачать руководство пользователя страница 5

Safety

2

Safety

2.1

General warning

WARNING!
Risk of injury

There is a risk of entrapment of fingers in the bed moving
parts.

Pay attention to your fingers.

IMPORTANT!

Make sure that there is nothing under, over, or near

the bed that can limit the movement of the bed or the
mattress support, such as furniture, window frames
and storage boxes.

Do not roll over the mains cable.

Do not bring mains cable into moving parts.

Disconnect the plug from the mains before moving

the bed.

Make sure that no cables (mains or from other

equipment) are jammed or damaged, when the bed is
used.

The bed must not be placed close to a direct heat

source (dire place, radiator etc) and not in direct
sunlight.

2.2

Labels and symbols on the product

Product label

The product label is placed on the frame of the bed and contains the
following information for your bed:

Type No

Serial No

Date of production

Manufacturer; Name and
address

Electrical data

Degree of protection.

Max user weight

Max working load

Standard code

Abbreviations

Iin — Current in

Uin — Voltage in

Int. — Intermittency

AC — Alternating Current

kg — kilogram

Max — maximum

min — minutes

A — Ampere

All symbols in this table might not be applicable for your
product. Please see product label and product.

Serial No

Address manufacturer

Production date

Max. user weight.

Max. safe working load.

Protection CLASS II equipment (double insulated).

Protection CLASS I.

Type B Applied Part.

Potential equalization.

Alternating current

Refer to the user manual.

After use. Waste disposal and recycling information.

This product complies with Directive 93/42/EEC
concerning medical devices. The launch date of this
product is stated in the CE declaration of conformity.

1581143-A

5

Содержание SB 400

Страница 1: ...Invacare SB 400 en Bed Service Manual DEALER Keep this manual The procedures in this manual MUST be performed by a qualified technician...

Страница 2: ...ission from Invacare Trademarks are identified by and All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted This bed has been tested and approved a...

Страница 3: ...th U profiles 11 3 5 Side Rails 12 3 5 1 Scala 2 side rail 12 3 5 2 Britt V and Line side rail 12 3 5 3 Verso ll side rail 12 3 6 Wiring 12 3 7 Mattress retainers 13 3 8 Mattress support extension 13...

Страница 4: ...user manual warnings are indicated by symbols The warning symbols are accompanied by a heading that indicates the severity of the danger WARNING Indicates a hazardous situation that could result in se...

Страница 5: ...uct label is placed on the frame of the bed and contains the following information for your bed Type No Serial No Date of production Manufacturer Name and address Electrical data Degree of protection...

Страница 6: ...nner 3 2 Mattress support Bed parts including mounted motors can be heavy Follow the instructions carefully and if possible be two persons Follow special instructions for two All accessories bed ends...

Страница 7: ...crews under the side tubes on each side of the mattress support 12 Connect all motor plugs to the control box see label on the control box for correct placement Make sure that all motor wires are plac...

Страница 8: ...the bed to the mains 3 2 3 Disassembly of mattress support 2 persons Recommended IMPORTANT Two persons are needed to carry out these instructions Person one in the head end of the bed Person two in t...

Страница 9: ...comes to a stop 9 Tip it gently onto the floor until it stands on the bed end attachments and set it aside 10 Move over to the lower half of the mattress support 11 Pull out the red buttonsL on each s...

Страница 10: ...spacers A and the ends of the shear arms onto the attachments B on the base frame 4 Attach the hi lo motor to the frame with the cotter pin Make sure the clamp is closing properly 5 Put back the cotte...

Страница 11: ...d Risks when using bed end with removable plate If the user is restless confused or has spasms Victoria Bed ends with removable plates must NOT be used Make sure that the side rails are combined with...

Страница 12: ...ntil they have all passed the locking pin D 4 Repeat steps 2 and 3 to mount the other side of the side rail Dismounting 1 Lower the side rail 2 Lift the lower bar in one end to make the locking pin D...

Страница 13: ...ets C on the outer laths Foot end Fold out the mattress retainer B 3 8 Mattress support extension For patients taller than two meters it is recommended to extend the bed with a mattress support extens...

Страница 14: ...the control unit B 2 Connect the cable to the control unit Charging the batteries 1 Connect the control unit to the mains The control unit must be connected to the mains for 24 hours prior to first u...

Страница 15: ...ed and undamaged q Electric plugs undamaged q Damaged coating repaired q Line side rail straps not frayed or cracked q Accessories correctly assembled and correct function q Rastofix fitting check fun...

Страница 16: ...equipment The washability of the components must be determined by authorized personnel If the components cannot withstand Jet base equipment or if it is unclear they have to be cleaned as non washable...

Страница 17: ...lace the control unit Mains indicator lights up but the motor is not running No relay sound is heard from the control unit Hand control is defective Replace the hand control Control unit is defective...

Страница 18: ...andard 41 kg Base frame with central brakes 46 kg 6 2 Dimensions side rails Side rail Measures l length h height Type m mounted r release button Line standard l 206 cm h 40 cm Full length aluminium m...

Страница 19: ...ss A or Class B Electromagnetic environment guidance RF emissions CISPR 11 partly Group I The medical bed uses RF energy only for its internal function Therefore its RF emissions are very low and are...

Страница 20: ...ec Mains power quality should be that of a typical commercial or hospital environment If the user of the medical bed requires continued operation during power mains interruptions it is recommended tha...

Страница 21: ...ed should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the medical bed b Over the frequency range 1...

Страница 22: ...Invacare SB 400 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 22 1581143 A...

Страница 23: ...Notes...

Страница 24: ...353 1 810 7085 ireland invacare com www invacare ie United Kingdom Invacare Limited Pencoed Technology Park Pencoed Bridgend CF35 5AQ Tel 44 0 1656 776 222 Fax 44 0 1656 776 220 uk invacare com www in...

Отзывы: