Invacare® SB® 400
4.4
Cleaning and disinfection
IMPORTANT!
Wrong fluids or methods can harm or damage you
product.
–
Follow instructions carefully for either Non-washable
or Washable components.
–
Never use corrosive fluids (alkalines, acid, cellulose
thinner, acetone etc)
–
Never use a solvent that changes the structure of the
plastic or dissolves the attached labels.
–
Always make sure that the bed is carefully dried before
taking it into use again.
4.4.1
Non-washable components
IMPORTANT!
Non-washable components cannot withstand washing
with Jet based equipment.
–
The washability of the components must be determined
by authorized personnel.
–
If the components cannot withstand Jet base
equipment or if it is unclear they have to be cleaned as
non-washable components.
Electrical components
Method:
Wipe off with a wet cloth or soft brush.
Max temp:
40 °C
Solvent/
chemicals:
Water
Metal components
Method:
Wipe off with a wet cloth or soft brush.
Max temp:
40 °C
Solvent/
chemicals:
Household detergent or soap and water, 6-8 pH
Wood (including textile straps on side rails, if existing)
Method:
Wipe off with a wet cloth or soft brush.
Max temp:
40 °C
Solvent/
chemicals:
Household detergent or soap and water, 6-8 pH
Textiles (including upholstery and mattresses)
Method:
See attached label on each product.