DANGER!
Risk of Injury or Death
While Invacare strives to produce the best oxygen
concentrator in the market today, this oxygen
concentrator can fail to produce oxygen due to
power failure or device malfunction.
– ALWAYS have a backup source of oxygen readily
available.
– In the event the concentrator fails to produce
oxygen, the concentrator will briefly alarm
signaling the patient to switch to their backup
source of oxygen. Refer to Troubleshooting for
more detail.
1.3 Indications For Use
To provide supplemental oxygen to patients with respiratory
disorders, by separating nitrogen from room air, by way of a
molecular sieve.
1.4 Description
The Invacare Platinum concentrator is used by patients with
respiratory disorders who require supplemental oxygen. The
device is not intended to sustain or support life.
The oxygen concentration level of the output gas ranges
from 87% to 95.6%. The oxygen is delivered to the patient
through the use of a nasal cannula.
The Invacare Platinum concentrator uses a molecular sieve
and pressure swing adsorption methodology to produce the
oxygen gas output. Ambient air enters the device, is filtered
and then compressed. This compressed air is then directed
toward one of two nitrogen adsorbing sieve beds.
Concentrated oxygen exits the opposite end of the active
sieve bed and is directed into an oxygen reservoir where it
is delivered to the patient.
The Invacare Platinum concentrator is capable of operation
by the patient in a home environment or in an institutional
environment. Device operates at a nominal 230 VAC/50
Hertz supply.
Service information will be available upon request to
qualified technical personnel only.
1.5 Contraindications
There are no known contraindications.
1195054-B
7
Содержание Platinum IRC9LXO2AWQ
Страница 46: ...Notes...
Страница 47: ...Notes...