Invacare® Platinum® Oxygen Concentrator
10 Technical Data
10.1 Specifications
Electrical Requirements:
230 VAC +10, -15% (253 VAC/195.5 VAC), 50 Hz
Rated Current Input:
2.3 A
Sound Pressure Level:
< 50 dBA weighted for 3 L/min flow
< 50 dBA weighted for 9 L/min flow
Sound Power Level:
< 50 dBA weighted for 9 L/min flow
Altitude:
Up to 1230 m (4,000 ft) above sea level without degradation of concentration levels.
Atmospheric Pressure Range: 101.33 kPa – 88.0 kPa
Oxygen Output Concentration
Levels:
92% minimum at 1 to 2 L/min
93% minimum at 3 to 7 L/min
91% minimum at 8 L/min
87% minimum at 9 L/min
Concentration levels achieved after initial warm-up period (approximately 30
minutes)
Maximum Outlet Pressure:
62.0 kPa ± 3.45 kPa (9 psi ± 0.5 psi)
Flow Range:
1 L to 9 L/min. Flowrates less than 1 L/min are not recommended.
Potential Obstruction Alert:
The concentrator detects a condition that may indicate a potential obstruction of the
output oxygen. Rapid audible beeping alert (this alert is deactivated when accessories
are connected) may be associated with flow setting of 0.5 L/min or less.
Power Consumption:
460 W Typical
36
1195054-B