7.4 Humidifier Cleaning and Thermic Disinfection
Clean and disinfect the oxygen humidifier daily to
reduce limestone deposits and eliminate possible
bacterial contamination. Follow the instructions
provided by the manufacturer. If none are provided,
follow these steps:
1.
Wash humidifier in soapy water and rinse with a
solution of ten parts water and one part vinegar.
2.
Rinse thoroughly with hot water.
3.
Air dry thoroughly.
To limit bacterial growth, air dry the humidifier
thoroughly after cleaning when not in use. Refer to
6.3 Setting Up the Concentrator, page 19
for use.
7.5 Cleaning and Disinfection Between Patients
WARNING!
Risk of Injury or Damage
To prevent injury from infection or damage to
concentrator:
– Only qualified personnel should perform
cleaning and disinfection of the oxygen
concentrator and accessories between patients.
Follow these instructions to eliminate possible
pathogen exchange between patients due to
contamination of components or accessories.
Preventive maintenance should also be performed
at this time if necessary.
1.
Dispose of and replace all patient side accessories not
suitable for multiple patient use, including but not
limited to:
•
Nasal Cannula and Tubing
•
Mask
•
Humidifier
2.
Perform maintenance procedures described in this
manual and items on the Preventive Maintenance
Checklist.
3.
Check concentrator for possible external damage or
signs that it may require service or repair.
4.
Ensure concentrator functions properly and all alarms
are in working order.
5.
Before repackaging and distribution to new patient,
ensure packaging contents contain the concentrator,
user manual, power cord, assembly instructions and
labels.
1195054-B