Invacare® Platinum® Oxygen Concentrator
Standards and Regulatory Listing:
Certified complying with:
IEC/EN 60601-1
IEC/EN 60601-1-2
IEC/EN 60601-1-6
IEC/EN 60601-1-11
ISO 80601-2-69
CE marked model:
IRC9LXO2AWQ
Electrical:
No extension cords.
Placement:
No closer than 30.5 cm (12 in) from any wall, furniture, draperies or similar surfaces to
assure sufficient air flow.
Avoid deep pile carpets and heaters, radiators or hot air registers.
Floor location only.
No confined spaces (Example: No closets).
Environment:
Smoke and soot-free
Tubing:
Recommended use up to 15.2 m (50 ft) high flow tubing with high flow cannula at
all flow rates.
Time of Operation:
Up to 24 hours per day
Recommended Storage and Shipping
Temperature:
-29°C to 65°C (-20°F to 150°F) at up to 95% relative humidity
Measurement uncertainty is included in the device specification. All conditions at STPD (standard temperature and
pressure dry).
38