Invacare® Action® Ampla™
1.
Dismount the transversal bar
D
with the form washer
H
from the vertical brace
I
by removing the adapter
sleeve
K
and washer
J
on both sides.
2.
Install the ball castors
G
to the transportation bearing
support
A
using bolts
E
and square nut
F
.
3.
Install the transportation bearing to the transversal bar
D
using the flange
B
and bolts
C
. Do not completely
tighten the bolts.
4.
Position the transportation bearing
A
in a distance of
35 mm measured from the end of the axis
H
.
5.
Tighten the bolts
C
.
6.
Install the second transportation bearing in the same
way on the other side.
7.
Reinstall the transversal bar with the transportation
bearing to the vertical braces.
C
= 16 Nm
E
= 7 Nm
K
= 45 Nm
24