R E M O T E I N T E R F A C E S
Device Specific Commands
ATS-515, -525 & ATS-615 & -625 Interface & Applications Manual
4-15
DSNS
Set the DUT sensor type.
DSNS n -- where n is 0-4
0 – no DUT sensor
1 – type T thermocouple
2 – type K thermocouple
DSNS?
Read the DUT sensor type.
DUTC
Set the device thermal constant.
DUTC nnn -- where nnn is nominally 100 but can range from 20 -
500.
NOTE
: Use a higher number for a higher mass device, and to
reduce the amount of overshoot. A lower number may cause some
overshoot, but may also reduce the transition time.
DUTC?
Read the device thermal constant.
DUTM
Turn DUT mode on or off.
DUTM 0 -- off (air control)
DUTM 1 -- on (DUT control)
DUTM?
Read DUT mode on/off state.
NOTE
: The DUT mode state also appears as a bit in AUXC?.
EROR?
Read the device-specific error register (16 bits).
bit 15 – reserved
bit 14 – no DUT sensor selected
bit 13 – reserved
bit 12 – BVRAM fault
bit 11 – NVRAM fault
bit 10 – No Line Sense
bit 9 -- flow sensor hardware error
bit 8 – reserved
bit 7 -- internal error
bit 6 – reserved
bit 5 – air sensor open
bit 4 -- low input air pressure
bit 3 -- low flow
bit 2 – reserved
bit 1 -- air open loop
bit 0 – overheat
FLOW
Turn the main nozzle air flow on or off.
FLOW 1 – on
FLOW 0 – off
FLWR?
Read the measured main nozzle air flow rate, in scfm.
NOTE
: This query and FLWR? are identical.
FLRL?
Read the measured main nozzle air flow rate, in liters/sec.
Command