S Y S T E M O P E R A T I O N
PROGRAM OPERATION “RAMP/CYCLE” Screen
ATS-515, -525 & ATS-615 & -625 Interface & Applications Manual
3-21
PROGRAM OPERATION “RAMP/CYCLE” Screen
PROGRAM
OPERATION
“RAMP/CYCLE”
Screen
Detailed
“RAMP/CYCLE”
Screen
Descriptions
Part
Function
ESC
program button
Press to exit the
PROGRAM TEST SETUP “RAMP/CYCLE”
Screen
and return to the previous screen.
ROW
program button
Press to cycle through Row # “0” through “12”.
• When the desired Row is selected, the
“SELECT”
program
button displays.
• Pressing the
“SELECT”
program button Ramps to the Setpoint
as specified in the highlighted Row.
FLOW
program button
Toggles air flow on/off.
If air flow is toggled off, the displays reads "Flow 0 scfm”.
CYCLE
program button
Press to begin cycling through the Setpoints. The System
sequentially runs to the Setpoint of each Row where Ramp °C is
set greater than zero.
When initiated:
• the “
CYCLE MODE”
message displays.
• the “
PAUSE”
and “
STOP”
program buttons displays. Each
respectively Pauses or Stops the cycling program.
SETUP
program button
Selects the
PROGRAM TEST SETUP
Screen
. This screen allows
the user to set 12 Program test setups, each with unique setpoints.
For greater detail, see