Screener PLUS Hardware Manual: Part Number 11811 Rev A, Maintenance : Clog Prevention and Management
Then click
Add…
which brings up the Add Rinse button.
Set the Rinse Station to
Buffer
and the time to
1
second and click
OK
.
Repeat these steps two more times to create a 3 second Buffer rinse configured as three 1 second rinses:
In this example, we’ve created a 3 second rinse after every 6 wells which will add about 50 seconds to a 96-well
plate run. However, it can sometimes be necessary to rinse more frequently, even as much as after every well for the
most challenging sample types. There is no hard-and-fast rule. Test different protocols and adjust the frequency and
duration of the rinse to match the needs of your particular sample and throughput.
Содержание iQue Screener PLUS
Страница 1: ...iQue ScreenerPLUS hardware manual December 2016...
Страница 2: ......
Страница 9: ...5 iQue Screener PLUS Hardware Manual Part Number 11811 Rev B Introduction Declaration of Conformity...
Страница 10: ...6 iQue Screener PLUS Hardware Manual Part Number 11811 Rev B Introduction THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 20: ...16 iQue Screener PLUS Hardware Manual Part Number 11811 Rev B Safety THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 60: ...56 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 62: ...58 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 66: ...62 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 78: ...74 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance 7 Fluidics Check 8 Priming the System...
Страница 90: ...86 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 94: ...90 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 96: ...92 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Special Topics THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 102: ...98 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Special Topics THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ......