![INTELLICYT iQue Screener PLUS Скачать руководство пользователя страница 41](http://html1.mh-extra.com/html/intellicyt/ique-screener-plus/ique-screener-plus_hardware-manual_2073912041.webp)
37
iQue
®
Screener PLUS Hardware Manual: Part Number 11811 Rev A, Operation: Daily QC
Daily QC
Overview
Running daily quality control is highly recommended. A daily QC check not only ensures that system performance
is within specification, but is also a useful diagnostic tool if issues arise. IntelliCyt makes the QC process easy by
providing reagents and pre-configured analysis templates. Daily QC should take less than 5 minutes to complete.
Procedure
The Daily QC test runs a set of Validation Beads and checks their performance against a set of standards. For each
fluorescence channel, the mean fluorescence and CV of the brightest peak is checked.
1.
Prepare Validation Beads by adding
2 drops
of
beads to
300 μL
Buffer Solution in a 1.5 mL microcentrifuge
tube.
Note: A single preparation of beads is sufficient for five days
of Daily QC runs and should last a week if stored in the dark at
2–8˚C.
2.
Vortex the bead sample and place the tube in the iQue Screener
PLUS tube holder.
3.
Ensure there is a Buffer cartridge in the Buffer slot in the fluid
trough holder.
4.
Add a plate to your current Daily QC Experiment.
5.
Run the plate (approximately 5 minutes).
Содержание iQue Screener PLUS
Страница 1: ...iQue ScreenerPLUS hardware manual December 2016...
Страница 2: ......
Страница 9: ...5 iQue Screener PLUS Hardware Manual Part Number 11811 Rev B Introduction Declaration of Conformity...
Страница 10: ...6 iQue Screener PLUS Hardware Manual Part Number 11811 Rev B Introduction THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 20: ...16 iQue Screener PLUS Hardware Manual Part Number 11811 Rev B Safety THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 60: ...56 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 62: ...58 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 66: ...62 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 78: ...74 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance 7 Fluidics Check 8 Priming the System...
Страница 90: ...86 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 94: ...90 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Maintenance THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 96: ...92 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Special Topics THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 102: ...98 iQue Screener PLUS Hardware Manual Part Number 11811 Rev A Special Topics THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 112: ......