CONFIDENTIAL and
PROPRIETARY
96-03996-00-01 revF
DCR-15-284
INOGEN ONE G3 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
Page 7 of 17
Flow
setting
Flow Rate
(ml +/- 10%)
180
min
nominal
max
min
nominal
max
min
nominal
max
1.0
210
18.9
21.0
23.1
9.5
10.5
13.6
7.6
8.4
9.2
2.0
420
37.8
42.0
46.2
18.9
21.0
27.2
15.1
16.8
18.5
3.0
630
56.7
63.0
69.3
28.4
31.5
40.8
22.7
25.2
27.7
4.0
840
75.6
84.0
92.4
37.8
42.0
54.4
30.2
33.6
37.0
5.0
1050
94.5
105.0
115.5 47.3
52.5
67.9
37.8
42.0
46.2
10 BPM
(+/- 3ml or +/- 10%)
20 BPM
(+/- 3ml or +/- 10%)
25 BPM
(+/- 3ml or +/- 10%)
DESIGN
NOTE
Bolus Volume is tuned to provide the correct bolus volume
when delivered through a nasal cannula such as the Salter
Labs 1600Q. If bolus volumes are measured without a cannula
or with a different type of nasal cannula, bolus volumes
will vary from values stated in the table above.
2.4.2.
Trigger sensitivity
The conserver will trigger when the negative pressure at the
internal sensor reaches -0.12 cm H
2
0 (+/- 20%). This low
trigger sensitivity allows for breath detection of very
shallow breathing, but may also result in the occasional
trigger due to motion of the cannula or motion of the
concentrator. The trigger sensitivity may vary
2.4.3.
Flow Setting Selection Relative to Physician
Prescription
Inogen has labeled each of the four settings to provide a
guideline
for matching the setting of the device to the
continuous flow prescription issued by the physician. Actual
correlation is dependent upon the patient’s breathing rates,
inspiratory tidal volume, and other physiologic factors.
Inogen suggests that each patient be titrated (a) while
sedentary, (b) while active or ambulatory.
2.4.4.
Use with the Inogen Satellite Conserver
The Inogen One G3 is not compatible with the Inogen One
Satellite Conserver
3.
SERVICING THE INOGEN ONE G3
3.1.
Maintenance by the Provider
3.1.1.
Authorized Repair Centers
The Inogen One G3 is only intended to be repaired by
authorized repair centers. If a repair is required, please
contact Inogen to locate your nearest authorized repair