3.1.6.
Oxygen Purity Check
The oxygen concentration can be checked using the Salter Labs
PrO2 Check ultrasonic oxygen analyzer. The concentrator should
be run for 10 minutes before measuring the oxygen concentration.
If the concentrator has not been used for more than 2 weeks, the
concentrator should be run for a minimum of 30 minutes to allow
for adjustments to the operating parameters prior to measuring
the oxygen concentration.
3.1.7.
Data Logging
Operating data from your Inogen One G3 concentrator are
periodically recorded to flash memory located within the device.
Data is stored for a period of approximately 6-12 months; new
data replaces the oldest data as it is collected.
Additionally, the device records errors and system information
at the time of the error which is useful in diagnostics.
3.1.7.1.
The following information is a list of recorded
parameters that can be used to evaluate the device and
its usage:
Real Time Clock
Life Clock
Flow Setting
Breaths per Minute
Low, Medium, and High Priority Errors
Ambient Pressure
System Temperature
Battery Status
Charging Status
3.1.8.
Data Download
To download data from the Inogen One G3, you will need a USB
mass storage device. If one of the following drives cannot be
obtained, the drive must be less than 2GB in capacity and have
a native file format of FAT32. The capacity of the drive and the
file system alone do not ensure compatibility with the Inogen
One G3.
Compatible USB drives include:
PNY Attache 2GB PN: P-FD2GBATT2-SF
SanDisk Cruzer Micro 2GB PN:SDCZ6-2048RB
Kingston Data Traveler G3 2GB PN:DTIG3/2GBZ