CONFIDENTIAL and
PROPRIETARY
96-03996-00-01 revF
DCR-15-284
INOGEN ONE G3 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
Page 5 of 17
2.
SETTING UP A PATIENT ON INOGEN ONE G3
2.1.
Indications for Use
The Inogen One G3 Oxygen Concentrator is used on a prescriptive
basis by patients requiring supplemental oxygen. It supplies a high
concentration of oxygen and is used with a nasal cannula to channel
oxygen from the concentrator to the patient. The Inogen One Oxygen
Concentrator may be used in a home, institution, vehicle and
various mobile environments.
NOTE
Availability of an alternate source of oxygen is
recommended in case of power outage. Several certifying
bodies for Home Health Care Providers require that back-up
oxygen be available to the patient. Supplemental oxygen
cylinders or extra Inogen One G3 Batteries may satisfy
these requirements.
CAUTION
Oxygen demand of some patients, particularly those with
high breathing rates and high flow settings, may exceed the
capabilities of the Inogen One G3. Inogen suggests that
each patient be titrated to assure that the Inogen One G3
is an appropriate solution for their needs.
2.2.
System Components
The following are standard components of the Inogen One system:
o
Inogen One G3 Oxygen Concentrator
o
8 Single Battery, BA-300
o
AC to DC Power Supply, BA-301
o
DC Power Cord
o
Carry Bag
o
Cannula
Patient Set-Up. To properly set up a patient on the Inogen One
G3 System, you may need to provide:
o
Pulse Oximeter (for titration, not included)
o
Extra cannulas (not included)
2.3.
Using the Inogen One G3
To quickly configure the Inogen One G3 for patient use:
1.
Slide the battery onto the bottom of the system until the
latch clicks into place.
2.
Connect AC power cord to the power supply.
3.
Connect the power supply to the Inogen One G3.