CONFIDENTIAL and
PROPRIETARY
96-03996-00-01 revF
DCR-15-284
INOGEN ONE G3 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
Page 14 of 17
3.3.
Expected Service Requirements
The following table is provided as an estimate only and
assumes typical environmental conditions for temperature,
humidity, and air pollution. Any smoking around the device
will severely shorten its life expectancy. Please refer to
product warranty coverage terms.
Service
Item
Estimated Service Requirement Frequency
Battery
500 full charge/discharge cycles (to
approximately 80% capacity)
Compressor
Function of flow setting and time
20,000 run hours @ flow setting 4
25,000 run hours @ flow setting 2
Oxygen
Sensor
No Service Requirement
Molecular
Sieve
18 to 36 months depending on usage
Valves
25,000 run hours
Intake
Filter
20,000 run hours
(without smoke or other abnormal
contaminants)
Cooling Fan
25,000 run hours
4.
INOGEN ONE G3 SYSTEM SPECIFICATIONS
INOGEN ONE G3 CONCENTRATOR
Dimensions:
Length
Width
Height
Cm In Cm In Cm In
No battery
22.2 8.75 7.6 3.0
18.4 7.25
8 cell
battery
22.2 8.75 7.6 3.0
21.0 8.25
16 cell
battery
22.2 8.75 7.6 3.0
23.5 9.25
Weight:
4.8 pounds (includes 8 cell battery)
Noise:
<42 dB (G3) or 39 db (G3HF) on flow setting 2
Warm-Up Time:
Approximately 2 minutes
Oxygen
Concentration:
90 + 6/-3%
Flow Control
Settings:
1,2,3,4,and 5 (G3HF only)
Power:
AC Power Supply
Input:100 to 240 VAC, 50 to 60 Hz, 1.0A to
0.5A