INOGEN ONE G3 OXYGEN CONCENTRATOR  

TECHNICAL MANUAL 

Page 15 of 17 

 

 
 

 

Output 19 VDC, 85 W 
 
DC Input: 13.5 to 19 VDC, 10A 
 
Rechargeable Battery: Voltage12.0 to 16.8 VDC, 

Battery Duration: 

Approximately 2 to 4 hours for 8 cell battery 
Approximately 4 to 8 hours for 16 cell battery 

Battery Charging 
Time: 

Approximately 2.25 to 4.25 hours for 8 cell 
battery 
Approximately 4.25 to 8.00 hours for 16 cell 
battery  

Environmental 
Ranges Intended 
for Use: 

Temperature: 40 - 104

o

F (4 - 40

o

C) 

Humidity: 0% to 95%, non-condensing 

 

Classifications 

 

Mode of Operation: 

Continuous Duty 

Type of Protection Against 
Electrical Shock: 

Class II 

Degree of Protection Against 
Electrical Shock: 

Type BF 
Not intended for cardiac 
application 

Degree of Protection Against 
Ingress of Water: 

Concentrator:    IP 20 
 w/carry bag:     IP 22 

Degree of Safety for Application 
in Presence of Anesthetic Gases: 

Not suitable for such 
applications 

 

5.

 

INOGEN ONE G3 ERROR CODE TABLE

 

Error Code 
Number 

Explanation

Possible Causes & 
Troubleshooting 
Instructions 

001 

Low voltage on power 
supply 

Replace the power supply or 
remove the battery to see 
if the concentrator was 
running on battery power 

004  

Pressure Error 

The concentrator was shut 
down due to a pressure 
error. Check that the sieve 
beds are installed 
properly. 

016 

System electric current 
out of specifications 

There may be a fault with 
the motor. If the error is 
persistent, return the 
concentrator for service 

Содержание IO-300

Страница 1: ...R TECHNICAL MANUAL Page 1 of 17 INOGEN ONE G3 OXYGEN CONCENTRATOR TECHNICAL MANUAL 96 03996 00 01 Revision F This Manual applies to the following Inogen Inc products Inogen One G3 Oxygen Concentrator model IO 300 Inogen One AC Power supply model BA 301 Battery model BA 300 Battery model BA 316 ...

Страница 2: ...2 SETTING UP A PATIENT ON INOGEN ONE G3 5 2 1 RECOMMENDATIONS FOR USE 5 2 2 SYSTEM COMPONENTS 5 2 3 USING THE INOGEN ONE G3 5 2 4 SELECTING THE PROPER FLOW SETTING 6 3 SERVICING THE INOGEN ONE G3 7 3 1 MAINTENANCE BY THE PROVIDER 7 3 2 MAINTENANCE BY THE PATIENT 11 3 3 EXPECTED SERVICE REQUIREMENTS 14 4 INOGEN ONE G3 SYSTEM SPECIFICATIONS 14 5 INOGEN ONE G3 ERROR CODE TABLE 15 6 ERROR RECALL 17 7 ...

Страница 3: ...are in relaying this information will not only enhance the user s experience with the Inogen One G3 but will also protect the patient prolong the life of the device and help you avoid unnecessary service calls and complaints from users 1 1 Caution and Warning Statements You will see Warnings and Cautions throughout this Technical Manual To ensure effective Oxygen Therapy and proper operation of th...

Страница 4: ...0 01 revF DCR 15 284 INOGEN ONE G3 OXYGEN CONCENTRATOR TECHNICAL MANUAL Page 4 of 17 information is not required to operate or maintain the Inogen One G3 In many cases warnings and cautions have been included in the Inogen One G3 User Manual ...

Страница 5: ...ra Inogen One G3 Batteries may satisfy these requirements CAUTION Oxygen demand of some patients particularly those with high breathing rates and high flow settings may exceed the capabilities of the Inogen One G3 Inogen suggests that each patient be titrated to assure that the Inogen One G3 is an appropriate solution for their needs 2 2 System Components The following are standard components of t...

Страница 6: ...nite stored volume of oxygen such as with compressed gas or liquid cryogenic systems The Inogen One G3 can provide oxygen to the patient as long as a source of electricity is available However because the oxygen is being produced as it is used supply of oxygen is rate limited The Inogen One G3 delivers up to 1050 ml min of 90 oxygen earlier models deliver up to 840 ml min of 90 oxygen At each flow...

Страница 7: ...t the internal sensor reaches 0 12 cm H20 20 This low trigger sensitivity allows for breath detection of very shallow breathing but may also result in the occasional trigger due to motion of the cannula or motion of the concentrator The trigger sensitivity may vary 2 4 3 Flow Setting Selection Relative to Physician Prescription Inogen has labeled each of the four settings to provide a guideline fo...

Страница 8: ...arranties Contact Inogen for information about receiving proper training and certification for service of the Inogen One G3 and accessories 3 1 3 Suggested Materials for Regular Maintenance To perform regular field maintenance by a technician on the Inogen One G3 System you may need o Replacement cannulas RP 128 o Replacement intake screens RP 300 and RP 301 o Replacement Inogen One G3 Batteries B...

Страница 9: ...w The Inogen One G3 includes a product filter conveniently located behind the removable cannula nozzle fitting Inogen suggests that this filter be replaced between patients To replace the Product Filter 1 Use the Cannula Barb Tool available from Inogen as RP 102 to access the product filter The tool has two prongs which mate with two indentations located on the surface of the metal cannula barb fi...

Страница 10: ...months new data replaces the oldest data as it is collected Additionally the device records errors and system information at the time of the error which is useful in diagnostics 3 1 7 1 The following information is a list of recorded parameters that can be used to evaluate the device and its usage Real Time Clock Life Clock Flow Setting Breaths per Minute Low Medium and High Priority Errors Ambien...

Страница 11: ...e Patient 3 2 1 Cannula Replacement The nasal cannula should be replaced on a regular basis A single lumen cannula of four to 25 feet in length must be used Inogen has certified its performance data with the Salter Laboratories 1600Q cannula CAUTION Do not use total cannula tubing length exceeding 25 feet with the Inogen One G3 unless proper saturation has been verified by a clinician 3 2 2 Intake...

Страница 12: ...er settings its battery life is extended Additionally at lower flow settings the concentrator does not generate as much heat and noise draw as much electric current from external power supplies and many system components do not wear as quickly Initial Battery State The Inogen One G3 batteries are shipped 40 to 100 charged The battery can be used immediately however we recommend the battery be char...

Страница 13: ...attery from the Inogen One G3 when it is not in use to avoid inadvertent discharge Leaving a battery attached to an idle Inogen One G3 for prolonged periods will result in battery damage that will severely shorten the expected life of the battery DESIGN NOTE When the concentrator is off but the battery installed the battery will continue to provide a small amount of power to the concentrator s mic...

Страница 14: ...20 000 run hours flow setting 4 25 000 run hours flow setting 2 Oxygen Sensor No Service Requirement Molecular Sieve 18 to 36 months depending on usage Valves 25 000 run hours Intake Filter 20 000 run hours without smoke or other abnormal contaminants Cooling Fan 25 000 run hours 4 INOGEN ONE G3 SYSTEM SPECIFICATIONS INOGEN ONE G3 CONCENTRATOR Dimensions Length Width Height Cm In Cm In Cm In No ba...

Страница 15: ...Against Electrical Shock Class II Degree of Protection Against Electrical Shock Type BF Not intended for cardiac application Degree of Protection Against Ingress of Water Concentrator IP 20 w carry bag IP 22 Degree of Safety for Application in Presence of Anesthetic Gases Not suitable for such applications 5 INOGEN ONE G3 ERROR CODE TABLE Error Code Number Explanation Possible Causes Troubleshooti...

Страница 16: ...e moved to a cooler location with improved ventilation Oxygen Low The Oxygen purity is 82 for 30 minutes The sieve beds may require servicing Oxygen Error The Oxygen purity is 50 for 10 minutes The concentrator has a malfunction such as a leak or a failed valve Sensor Fail The oxygen sensor is returning false readings If the problem persists return the concentrator for service Battery Error Check ...

Страница 17: ...LL Certain errors can be recalled so that the device displays the last error that occurred The last error stored in memory may be accessed by holding the Plus button for 5 seconds while the unit is in standby mode All errors will be stored except when the unit is operating in special admin mode no breath detect battery low and Error 1 7 CONTACTS FOR MORE INFORMATION Inogen Website Patient Issues h...

Отзывы: