The number of cycles that the battery will last is highly 
dependent upon the temperature at which the battery is 
charged.  Inogen recommends that batteries not be charged 
at room temperatures exceeding 75

o

F (24

o

C).   

 

DESIGN 

NOTE

 

The BA-300 and BA-316 will typically achieve 500 
charge/discharge cycles while retaining 80% of their original 
capacity if proper battery care is taken. 

 

Battery Time Remaining Clock 

The Inogen One G3 continuously displays battery time 
remaining. It may be necessary to inform the user that this 
displayed time is 

only an estimate

, and the actual time 

remaining may vary from this value. 
 
To avoid running out of battery power unexpectedly, users 
should regularly monitor the displayed battery time 
remaining and/or carry a back-up power supply (extra 
charged battery or AC Power Supply). 

 

Storage 

Instruct patients to remove the battery from the Inogen One 
G3 when it is not in use to avoid inadvertent discharge.  
Leaving a battery attached to an idle Inogen One G3 for 
prolonged periods will result in battery damage that will 
severely shorten the expected life of the battery. 
 

DESIGN 

NOTE

 

When the concentrator is off but the battery installed, the 
battery will continue to provide a small amount of power to 
the concentrator’s microprocessor.  This power draw will 
empty a full battery in approximately 2-3 weeks.   

 

CAUTION

 

Leaving a Battery in the Inogen One G3 while the device is 
unused will irreversibly damage the battery.  After such 
storage, the battery may not be able to recharge or its 
life cycle and/or capacity will be greatly diminished. A 
full battery can be damaged in as little as 20 days if left 
in the concentrator while not plugged in.  

 

Instruct users to avoid storing the Inogen One G3 battery 
in extreme temperatures, below -4

o

F (-20

o

C) or above 140

o

(60

o

C), for any amount of time.  They should avoid leaving 

batteries in automobiles, where these temperatures can be 
regularly exceeded.  Storage of the Inogen One G3 battery 
in a cool, dry location will help to maximize the longevity 
of the battery. 

Содержание IO-300

Страница 1: ...R TECHNICAL MANUAL Page 1 of 17 INOGEN ONE G3 OXYGEN CONCENTRATOR TECHNICAL MANUAL 96 03996 00 01 Revision F This Manual applies to the following Inogen Inc products Inogen One G3 Oxygen Concentrator model IO 300 Inogen One AC Power supply model BA 301 Battery model BA 300 Battery model BA 316 ...

Страница 2: ...2 SETTING UP A PATIENT ON INOGEN ONE G3 5 2 1 RECOMMENDATIONS FOR USE 5 2 2 SYSTEM COMPONENTS 5 2 3 USING THE INOGEN ONE G3 5 2 4 SELECTING THE PROPER FLOW SETTING 6 3 SERVICING THE INOGEN ONE G3 7 3 1 MAINTENANCE BY THE PROVIDER 7 3 2 MAINTENANCE BY THE PATIENT 11 3 3 EXPECTED SERVICE REQUIREMENTS 14 4 INOGEN ONE G3 SYSTEM SPECIFICATIONS 14 5 INOGEN ONE G3 ERROR CODE TABLE 15 6 ERROR RECALL 17 7 ...

Страница 3: ...are in relaying this information will not only enhance the user s experience with the Inogen One G3 but will also protect the patient prolong the life of the device and help you avoid unnecessary service calls and complaints from users 1 1 Caution and Warning Statements You will see Warnings and Cautions throughout this Technical Manual To ensure effective Oxygen Therapy and proper operation of th...

Страница 4: ...0 01 revF DCR 15 284 INOGEN ONE G3 OXYGEN CONCENTRATOR TECHNICAL MANUAL Page 4 of 17 information is not required to operate or maintain the Inogen One G3 In many cases warnings and cautions have been included in the Inogen One G3 User Manual ...

Страница 5: ...ra Inogen One G3 Batteries may satisfy these requirements CAUTION Oxygen demand of some patients particularly those with high breathing rates and high flow settings may exceed the capabilities of the Inogen One G3 Inogen suggests that each patient be titrated to assure that the Inogen One G3 is an appropriate solution for their needs 2 2 System Components The following are standard components of t...

Страница 6: ...nite stored volume of oxygen such as with compressed gas or liquid cryogenic systems The Inogen One G3 can provide oxygen to the patient as long as a source of electricity is available However because the oxygen is being produced as it is used supply of oxygen is rate limited The Inogen One G3 delivers up to 1050 ml min of 90 oxygen earlier models deliver up to 840 ml min of 90 oxygen At each flow...

Страница 7: ...t the internal sensor reaches 0 12 cm H20 20 This low trigger sensitivity allows for breath detection of very shallow breathing but may also result in the occasional trigger due to motion of the cannula or motion of the concentrator The trigger sensitivity may vary 2 4 3 Flow Setting Selection Relative to Physician Prescription Inogen has labeled each of the four settings to provide a guideline fo...

Страница 8: ...arranties Contact Inogen for information about receiving proper training and certification for service of the Inogen One G3 and accessories 3 1 3 Suggested Materials for Regular Maintenance To perform regular field maintenance by a technician on the Inogen One G3 System you may need o Replacement cannulas RP 128 o Replacement intake screens RP 300 and RP 301 o Replacement Inogen One G3 Batteries B...

Страница 9: ...w The Inogen One G3 includes a product filter conveniently located behind the removable cannula nozzle fitting Inogen suggests that this filter be replaced between patients To replace the Product Filter 1 Use the Cannula Barb Tool available from Inogen as RP 102 to access the product filter The tool has two prongs which mate with two indentations located on the surface of the metal cannula barb fi...

Страница 10: ...months new data replaces the oldest data as it is collected Additionally the device records errors and system information at the time of the error which is useful in diagnostics 3 1 7 1 The following information is a list of recorded parameters that can be used to evaluate the device and its usage Real Time Clock Life Clock Flow Setting Breaths per Minute Low Medium and High Priority Errors Ambien...

Страница 11: ...e Patient 3 2 1 Cannula Replacement The nasal cannula should be replaced on a regular basis A single lumen cannula of four to 25 feet in length must be used Inogen has certified its performance data with the Salter Laboratories 1600Q cannula CAUTION Do not use total cannula tubing length exceeding 25 feet with the Inogen One G3 unless proper saturation has been verified by a clinician 3 2 2 Intake...

Страница 12: ...er settings its battery life is extended Additionally at lower flow settings the concentrator does not generate as much heat and noise draw as much electric current from external power supplies and many system components do not wear as quickly Initial Battery State The Inogen One G3 batteries are shipped 40 to 100 charged The battery can be used immediately however we recommend the battery be char...

Страница 13: ...attery from the Inogen One G3 when it is not in use to avoid inadvertent discharge Leaving a battery attached to an idle Inogen One G3 for prolonged periods will result in battery damage that will severely shorten the expected life of the battery DESIGN NOTE When the concentrator is off but the battery installed the battery will continue to provide a small amount of power to the concentrator s mic...

Страница 14: ...20 000 run hours flow setting 4 25 000 run hours flow setting 2 Oxygen Sensor No Service Requirement Molecular Sieve 18 to 36 months depending on usage Valves 25 000 run hours Intake Filter 20 000 run hours without smoke or other abnormal contaminants Cooling Fan 25 000 run hours 4 INOGEN ONE G3 SYSTEM SPECIFICATIONS INOGEN ONE G3 CONCENTRATOR Dimensions Length Width Height Cm In Cm In Cm In No ba...

Страница 15: ...Against Electrical Shock Class II Degree of Protection Against Electrical Shock Type BF Not intended for cardiac application Degree of Protection Against Ingress of Water Concentrator IP 20 w carry bag IP 22 Degree of Safety for Application in Presence of Anesthetic Gases Not suitable for such applications 5 INOGEN ONE G3 ERROR CODE TABLE Error Code Number Explanation Possible Causes Troubleshooti...

Страница 16: ...e moved to a cooler location with improved ventilation Oxygen Low The Oxygen purity is 82 for 30 minutes The sieve beds may require servicing Oxygen Error The Oxygen purity is 50 for 10 minutes The concentrator has a malfunction such as a leak or a failed valve Sensor Fail The oxygen sensor is returning false readings If the problem persists return the concentrator for service Battery Error Check ...

Страница 17: ...LL Certain errors can be recalled so that the device displays the last error that occurred The last error stored in memory may be accessed by holding the Plus button for 5 seconds while the unit is in standby mode All errors will be stored except when the unit is operating in special admin mode no breath detect battery low and Error 1 7 CONTACTS FOR MORE INFORMATION Inogen Website Patient Issues h...

Отзывы: