Innovo® CMS 50F PLUS User Manual
31
Guidance and manufacture’s declaration – electromagnetic
immunity –
for EQUIPMENT and SYSTEMS that are not
LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic
immunity
The
CMS50F PLUS Pulse Oximeter
is intended for use in the
electromagnetic environment specified below. The user should
assure that it is used in such an environment.
Immunity
test
IEC 60601
test level
Comp
liance
level
Electromagnetic
environment -
guidance
Conducted
RF
IEC
61000-4-6
Radiated
RF
IEC
61000-4-3
3 V
rms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
3 V
rms
3 V/m
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the
CMS50F
PLUS Pulse Oximeter,
including cables, than
the
recommended
separation
distance
calculated
from
the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance