Innovo® CMS 50F PLUS User Manual
15
If the alarm function is on, the device will provide
medium-priority alarm signal when the data of SpO
2
or pulse
rate falls beyond set limit. Intermittent alarm will occur and
the measurement will be displayed in yellow font.
Medium priority indicating that prompt operator response is
required.
The alarm state setting
PUSH the
button
to select "Alarm". HOLD the
button
to set
alarm on or off. Select "on" to turn on the alarms and "off" to turn
off the alarms.
Pulse sound indication setting
PUSH the
button
to select "Pulse Sound". HOLD the
button
to
set Pulse Sound (heart beat) alarm to "on" or "off".
Exit the Alarm settings
PUSH the
button
to select "EXIT". HOLD the
button
to exit the
Alarm Settings Menu.
3) Clock setting
On the main menu interface, PUSH the
button
to select "Clock".
HOLD the
button
to enter the clock setting interface.
Figure 9. Clock Setting Menu
A) When entering the clock setting menu, “Set time” will always
be set to “NO” to prevent unexpected changes to the time as
shown in Figure 9. If you want to change the time, HOLD the
button
to change “Set time” to “YES”.