Innovo® CMS 50F PLUS User Manual
2
with the requirements of IEC 60601-1, or the device might be
damaged.
Please do not use the device while charging.
The device should only be used with the Acc U Rate
®
SnugFit probe. Using the Contec CMS 50F probe might
damage the oximeter.
1.3 Attention
Keep the oximeter away from dust, vibration, corrosive
substances, explosive materials, high temperature and moisture.
If the oximeter gets wet, stop operating it immediately.
When it is carried from cold environment to warm or humid
environment, please do not use it immediately.
DO NOT operate keys on front panel with sharp materials.
High temperature or high pressure steam disinfection of the
oximeter is not permitted. Refer to User Manual in the relative
chapter (7.1) for instructions of cleaning and disinfection.
Do not immerse the oximeter in liquid. To clean it, please wipe
its surface with medical alcohol with soft material. Do not spray
any liquid on the device directly.
When cleaning the device with water, the temperature should
be lower than 60
°
C.
The performance of the device is affected if the fingers are too
small or cold. As such, please clip and insert the thumb or middle
finger as deeply as possible into the probe.
The pulse oximeter can be used for adult or infant. However, a
special probe is required for infant. Please contact the
manufacturer for more details.
During measurement, the data is updated on average, every 5
seconds. However, this might change depending on individual.
Please read the measure value only when the amplitude of the
waveform is equal and steady. At this point, the measured value