27
Declaration and guide of manufacturer – electromagnetic immunity
INN-019 is designed to work in the electromagnetic environment specified below. The
customer or the user of the appliance should ensure that it is used in this environment.
Portable and mobile RF communications equipment should not be used closer to any part
of INN-019, including cables, than the recommended separation distance calculated with
the equation applicable to the transmitter frequency
Immunity test
Test level of IEC 60601
Conformity
level
Recommended
separation distance d
RF Irradiate
IEC 61000-4-3
3 V/m and 10 V/m
from 80 MHz to 2,7 GH
3 V/m and
10 V/m
d= 30 cm
mmunity to proximity
fields by wireless
RF communication
devices
IEC 61000-4-3
TETRA 400
380-390 MHz
27 V/m
27 V/m
d= 30 cm
GMRS 460 FRS 460
430-170 MHz
28V/m
28 V/m
LTE Band 13, 17
704-787 MHz
9V/m
9 V/m
GSM 800/900, TETRA 800,
iDEN 820, CDMA 850, LTE
Band 5 800 960 MHz
28V/m
28 V/m
GSM 1800; CDMA 1900;
GSM 1900; DECT; LTE Band
5 1700-1990 MHz
28V/m
28 V/m
Bluetooth, WLAN, 802.11
b/g/n, RIFD 2450, LTE Band
70 2400-2570 MHz
28V/m
28 V/m
WLAN 802.11 a/n
5100-5800 MHz
9V/m
9 V/m
GB