View950M Multi-Scan
INNO Instrument Inc. Page 48 of 55
<Run port 4>
<Proceed port 4>
<Complete port 4>
5.
Complete E-Cal on all ports
- End E-Cal for all ports or E-Cal for the desired port, then click the
Close button in the E-Cal Window.
- Calibration icon of each port is changed to green
.
4.2.
M
ULTI
T
ARGET TO
S
INGLE
L
IMIT
P
ROCEDURE
The VIEW950M is designed for the antenna production process, and VSWR of each band is measured
in most antenna manufacturing process to judge the quality of the product. If the frequency is within
a certain range, it is not in good condition. If the VSWR value is less than the predetermined value, it
is defective. If the VSWR value is less than the predetermined value, it is processed as non-conforming
product.