View950M Multi-Scan
INNO Instrument Inc. Page 46 of 55
- Press 'Apply' to display the message 'Port 1 is in the Cal On state.'
- At the same time, the Apply button disappears from the lower
button array.
5.
Proceed Calibration on the entire ports
- Proceed OSL calibration in the same manner with the rest of the ports
<
Complete port 2>
<
Complete port 3>
<
Complete port 4
- After completing the port calibration, click the Close button to
switch to the measurement screen.
- Calibration icon of each port is changed to green
4.1.2.
E-C
AL
C
ALIBRATION
1.
Proceed E-Cal – 1 port
- Connect USB port of E-Cal to USB port of VIEW950M as shown on
the left
- Connect the RF port of E-Cal to port 1 of VIEW950M
- Click the 'Run' button at the bottom