INNO Instrument Inc. Page 44 of 55
4.1.1.
OSL
C
ALIBRATION
1.
OSL – Open Calibration
- After selecting the OSL method, click the 'Start' button and a
calibration window will appear asking you to connect 'Open' to
the selected port as shown on the left.
- Connect Open to the selected port of the mechanical calibration
kit.
- After connecting Open to the port, click the Open Icon as shown
on the left.
- Click to open calibration as below.
<
Connection >
<
Progress >
<
Completion >
2.
OSL – Short Calibration
- When Open is completed, a screen for connection of Short is
displayed.
- Remove Open and connect Short.
- After connecting short to the port, click the Short Icon as shown
on the left.
- After clicking, short calibration will proceed as below.