View950M Multi-Scan
INNO Instrument Inc. Page 43 of 55
4.1.
C
ALIBRATION
P
ROCEDURE
The VIEW950M supports two types of calibration. It is an E-Cal method using an OSL method and an
Electronic Calibration Kit using a mechanical calibration kit. The procedure for the VIEW950M differs
according to the two methods. Calibration Proceed as follows.
1.
Setup equipment including Stimulus, Scale
- When changing some setting parameters, you have to perform calibration again.
- Therefore, before proceeding with the calibration, check the frequency setting of each port.
2.
Select Calibration Method – OSL or E-Cal
- Select 'OSL' when performing calibration with
mechanical calibration kit.
- Select 'E-Cal' when performing calibration with
Electronic Calibration kit
- After selecting, click 'Start' button.
<OSL Start screen>
<E-Cal Start screen>
Notice
-
In case of E-Cal, if USB of VIEW950 and E-Cal is not connected, Calibration Window can not enter.
Содержание View 950M
Страница 1: ...Revision 1 0...
Страница 4: ...View950M Multi Scan INNO Instrument Inc Page 4 of 55 1 GENERAL INFORMATION...
Страница 7: ...View950M Multi Scan INNO Instrument Inc Page 7 of 55...
Страница 8: ...View950M Multi Scan INNO Instrument Inc Page 8 of 55 2 INSTRUMENT OVERVIEW...
Страница 12: ...View950M Multi Scan INNO Instrument Inc Page 12 of 55 3 MENU DESCRIPTIONS...
Страница 42: ...View950M Multi Scan INNO Instrument Inc Page 42 of 55 4 TEST METHOD...
Страница 52: ...View950M Multi Scan INNO Instrument Inc Page 52 of 55 5 WARRANTY INFORMATION...
Страница 55: ...www innoinstrument com...