Transpector CPM Operating Manual
1.7.1.4 Fuse Rating
3.15 A @ 250 V
~
Interrupt Current 35 A @ 250 V
~
Type T (5 x 20 mm)
1.7.1.5 Overvoltage Category
Overvoltage Category II (per EN61010-1:2010)
Short Term: 1440 V < 2 seconds
Long Term: 490 V < 5 seconds
1.7.1.6 Electrical Connections
One of the following:
110 V
~
, three-pronged, grounded plug
230 V
~
, European style, two-pronged plug with ground contact
NOTE:
Do not replace power cord with an inadequately rated cord. Cord must be
rated 10 A or higher.
1.8 Nitrogen Purge Gas (Corrosive System Only)
The corrosive service Turbo Molecular Pump requires 10 to 25 sccm nitrogen
purge gas flow through its bearings to protect the bearings from corrosion and loss
of lubricant by evaporation. The CPM nitrogen regulator range is
10–15 psi (gauge). It is set for 10 sccm flow at the factory.
CAUTION
The corrosive service Turbo Molecular Pump (CVD/Etch),
requires nitrogen purge gas flow whenever the system is
operational.
Normally, dry nitrogen is supplied at the acceptable pressure range for the air
pressure. (See
). In this manner, dry nitrogen can be
supplied directly to the solenoid valve block and to the pressure regulator via the
supplied 6.35 mm (1/4 in.) tee. The regulator supplies dry nitrogen at a reduced
pressure for the purge.
Содержание Transpector CPM
Страница 1: ...O P E R A T I N G M A N U A L PN 074 641 P1B Cover Page Transpector CPM Compact Process Monitor ...
Страница 3: ...Declaration Of Conformity Page 1 ...
Страница 74: ...5 12 Transpector CPM Operating Manual Figure 5 6 Air mass spectrum N N2 O2 Ar O H2O ...
Страница 171: ...10 18 Transpector CPM Operating Manual Figure 10 29 Add mass ...