INFICON
Operation | 6
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
65 / 152
6.3 Settings for the measurements
6.3.1 Select operation mode
ü
Operator
or
Supervisor
rights
1
> Operation mode
2
Select between "Vacuum", "Sniffer / SL200" and "Sniffer / SL3000".
3
Save
.
6.3.2 Select gas
DANGER
Danger from a Hydrogen explosion
Hydrogen can explode in combination with oxygen. The allowable composition of
venal gas mixtures can be read in the safety data sheets of the respective
manufacturers.
► Ensure that the share of hydrogen does not exceed the described concentration.
ü
Supervisor
rights
1
> Setup > Measurement > Mass
2
Make a selection from:
"Hydrogen" H
2
(2 amu)
"Helium"
4
He (4 amu)
"Mass 3"
3
He (3 amu)
3
Save
.
6.3.3 Setting setpoints
You can set the leak rate for the setpoint to 1, 2, 3 and 4 separately.
When the setpoints are exceeded:
• If the setpoint 1 or 2 is exceeded, the measurement line in the measurement
window changes color.
• The setpoint relay of the digital output switches, see also "I/O module [
130]" or
the interface description.
Also setpoint 1 defines the trigger point for the different alarms, see also "Change
58]". It is also the limit for I•ZERO, see “Set and use the function
69]“.
ü
Operator
or
Supervisor
rights
1
> Setpoints
Содержание Fab UL6000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 125 152 ...
Страница 145: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 145 152 11 2 CE Declaration of Conformity ...
Страница 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 ...
Страница 147: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 147 152 11 3 RoHS ...
Страница 150: ...Index INFICON 150 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 Factory setting 70 ...
Страница 151: ......
Страница 152: ......