8 | Cleaning and maintenance
INFICON
122 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
8.9 Maintenance plan
Assembly
Maintenance work
Operating hours / years
Servic
e level
Spare part
number
1500
4000
8000
16000
24000
1/4
1
2
3
4
Vacuum system
Forevacuum
pump NeoDry
Check performance
X
X
X
X
II
Replace pump if
necessary
X
3
X
3
X
X
3
III
200009420
200009420R
TMP SplitFlow
80
Changing the oil wick
cartridge
X
3
II
200003801
Change bearings and
replace oil wick cartridge
X
2
III
Valve block
Clean valves
X
3
X
X
X
III
Replace valve seals
X
3
X
X
X
III
200007630
Dismantle and clean the
valve block
X
3
X
X
3
III
Flood and
purge gas line
Renew filter
X
3
X
1
X
1
X
1
II
200000683
200001095
Electrical
Fan
subassemblies
Blow out fan chassis and
TMP
X
3
X
1
X
1
X
1
X
1
II
Check air filter kit and
replace if necessary
X
3
X
1
X
1
X
1
X
1
I
200007656
Subassemblies and accessories
Sniffer line
SL3000
Renew filter
X
3
- Perform sniffer operation every
1000 hours
I
200001116
20003500
Calibration
leak
Recalibration /
replacement
X
2
- Perform depending on the expiration
date
III
14030
Maintenance
schedule legend
• I Customer or higher level
• II Customer with instruction or higher level
• III INFICON Service Technician
• X Carry out maintenance as per operating hours or duration
• X
1
Maintenance after operating hours, not after duration
• X
2
Maintenance by duration, not by operating hours
• X
3
Dependent on environmental influences, operating conditions, contamination
and application process
Содержание Fab UL6000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 125 152 ...
Страница 145: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 145 152 11 2 CE Declaration of Conformity ...
Страница 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 ...
Страница 147: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 147 152 11 3 RoHS ...
Страница 150: ...Index INFICON 150 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 Factory setting 70 ...
Страница 151: ......
Страница 152: ......