7 | Warning and error messages
INFICON
112 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
Type
Notification
Possible sources of
error
Remedy
W630
Calibration request
Operation mode or
mass has changed
• Perform a calibration
HYDRO•S has been
switched on (if function
available)
• Perform a calibration
Auto Leak Test is
active and testing time
has changed
• Perform a calibration
A new SL3000 sniffer
line has been
connected (if function
available)
• Perform a calibration
Runtime over 30
minutes and
temperature change by
5 Kelvin since last
calibration
• Perform a calibration
Runtime over 30
minutes and no
calibration for 24 hours
• Perform a calibration
W640
Auto Leak Test: Too many
consecutive tests failed
Leak rate setpoint 1
does not match the test
object
• Check and change the leak rate setpoint
1 if necessary
Test chamber leaky
• Use another test chamber, if possible
W670
Calibration error (emission
collapsed)
During calibration the
emission collapsed
• Check if the message occurs again when
you perform a new calibration
• Contact customer service
W700
Preamplifier temperature is
too low
Ambient temperature is
too low
• Increase the temperature in the
environment where the device is located
Temperature sensor is
defective
• Contact customer service
W702
Preamplifier temperature is
too high
The ambient
temperature is too high
• Reduce the temperature in the
environment where the device is located
Air filter dirty
• Clean the air vents or replace the filter
plates
Fan blocked or
defective
• Contact customer service
Содержание Fab UL6000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 125 152 ...
Страница 145: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 145 152 11 2 CE Declaration of Conformity ...
Страница 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 ...
Страница 147: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 147 152 11 3 RoHS ...
Страница 150: ...Index INFICON 150 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 Factory setting 70 ...
Страница 151: ......
Страница 152: ......