6 | Operation
INFICON
94 / 152
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
6.15 Logging off from the device
1
Press your name, which is displayed in the upper left corner of the display, or
select
> User accounts.
ð
The window "User accounts" opens. See also "Modify Personal Settings
54]".
2
You log off from the device via the button "Log off".
ð
The login window opens.
6.16 Switching off the device
NOTICE
Turbo molecular pumps can be damaged
The running turbo molecular pumps can be damaged by jerky movements or foreign
objects.
► Before switching off the device, vent the inlet to ensure that the TMP2 (booster
pump) is slowed down.
► After venting the inlet, avoid jerky movements or vibrations of the device during
operation and until 2 minutes after switching off.
► If you do not vent the inlet, avoid jerky movements or vibrations of the device during
operation and until 6 minutes after switching off.
► Prevent foreign objects from entering the inlet of the device.
You can switch off the device with the power supply switch at any time. The
parameters set in the device remain saved.
Содержание Fab UL6000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 125 152 ...
Страница 145: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 145 152 11 2 CE Declaration of Conformity ...
Страница 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 ...
Страница 147: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 147 152 11 3 RoHS ...
Страница 150: ...Index INFICON 150 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 Factory setting 70 ...
Страница 151: ......
Страница 152: ......