INFICON
Operation | 6
UL6000-Fab,-PLUS-Operating-instructions-iinc75en1-02-(2207)
63 / 152
3
Save
.
6.2.14 Measure internal calibration leak after run-up
After activating this option, the function "Measure internal calibration leak after run-up"
is automatically executed after run-up. This can be used to check that the leak
detector is functioning correctly.
The function is only available in the operation mode Vacuum with test gas Helium
(mass 4). For optimum measurement accuracy, calibration should be performed with
the device at operating temperature.
ü
Supervisor
rights
1
> Setup > Measurement > Miscellaneous
2
If necessary, change the setting of the "Measure internal test leak after run-up"
field.
ð
Option activated: Measurement activated
ð
Option deactivated: No measurement (factory setting)
3
Save
.
6.2.15 Use Favorites
To shorten the time spent searching menus for frequently used functions, create freely
assignable menu keys in the "Favorites" window.
Use Favorites:
►
Press the icon with your permission
,
,
and then the “Favorites” button
or choose alternatively
►
> User accounts > Favorites
ð
The Favorites window with 9 buttons is displayed. Assigned buttons can be used
to quickly call up desired functions.
Create Favorites:
1
In the Favorites window, press an empty button.
2
Select the desired menu name from the displayed list overview and confirm
your selection with
.
ð
Alternatively, you can press
in the "Favorites" window, mark an empty
entry and select the desired menu name from the list overview using the
shown.
Change or delete favorites:
1
In the Favorites window, press
.
ð
This opens the “Manage” window in which you can tap the desired entry.
Содержание Fab UL6000 Fab
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 125: ...INFICON Decommissioning 9 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 125 152 ...
Страница 145: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 145 152 11 2 CE Declaration of Conformity ...
Страница 146: ...11 Appendix INFICON 146 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 ...
Страница 147: ...INFICON Appendix 11 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 147 152 11 3 RoHS ...
Страница 150: ...Index INFICON 150 152 UL6000 Fab PLUS Operating instructions iinc75en1 02 2207 Factory setting 70 ...
Страница 151: ......
Страница 152: ......